Technical data
03/2018
1009909-05
Operating instructions – 55/70
5
Technical data
5.1
Power, connection rating and weight
MEDICAL_PLUS
Power consumption rating:
2800 W
Rated frequency:
50 Hz
Rated voltage:
230-240 V ~
Power consumption:
14 A
Rated fuse:
16 A
Connection type:
Landline connection (3.00 m)
Noise level 1m away from device:
57.0 dB(A)
Noise level inside the device::
< 61 dB(A)
Classification
by the type of protection against electric shock:
by the degree of protection against electric shock:
Protection class l
Type BF application part
Protection class:
I
Protection type:
IP20
Empty weight (unfilled):
ca. 168 kg
Weight:
ca. 472 kg
No continuous operation:
on < = 45 min
off > 1 min
Essential performance features
NOTE
The essential performance features of the device are carried
out within the following limits:
Lower limit
Massage pressure: 0.50
Temperature: 20°C
Application time: 5 Min.
Upper limit
Massage pressure: 4.00
Temperature: 40°C
Application time: 45 Min.