MEDICAL_PLUS: EMC information
7.1.1
Table 201: Guidelines and manufacturer's
declaration – Electromagnetic emissions
The MEDICAL_PLUS massage device has been designed for use in the elec-
tromagnetic environment specified below. The device operator must ensure
that it is used such an environment.
Emission measurements
Conformity
Electromagnetic environment – Guidelines
HF emissions as per CISPR 11
Group 1
MEDICAL_PLUS uses HF power solely for internal
functions. Therefore the HF emissions are very low and
it is highly unlikely that it will cause any interference to
adjacent electronic equipment.
HF emissions as per CISPR 11
Class B
MEDICAL_PLUS has been designed for use in all de-
vices, including those in residential areas, and it has
been determined that it can be used for residential pur-
poses as soon as it is connected up to a public mains
supply that also supplies buildings.
Harmonics
IEC 61000-3-2
Class A
Voltage fluctuations / Flicker
IEC 61000-3-11
Fulfilled
The obligations listed in Sub-section 6.2.2 of EN 61000-
3-11:2000 must be taken into consideration. In particu-
lar, the 0.14 ohm line impedance must be guaranteed
for pulse massaging.