Operation
46/70
3.9.2
Variable settings
Massage time, massage pressure, speed and body size settings
Operating
button
Settings /
Display
Remarks
START / STOP button
Massage time display
The device switches on.
After starting:
The variable settings can be defined
separately for the specific massage
areas and the the massage types,
e.g.:
1.
Step (massage area):
Whole body button
2.
Step (massage type):
Parallel massage
Altering the massage
time,
e.g. from
to
Press the massage time button and
hold down for approx. 1 s. The display
will flash. Press the plus or minus
buttons to set up the value that you
want.
Basic setting:
Changes can only be made using the
service card in the default settings
mode.
Factory setting: 30 minutes
Minimum and maximum massage
times in minutes:
5.00 to 45.00 in minutes
Altering the
massage pressure,
e. g. from
to
Press the massage pressure button
and hold down for approx. 1 s. The
display will flash. Press the plus or
minus buttons to set up the value that
you want.
Basic setting:
Changes can only be made using the
service card in the default settings
mode.
Factory setting:1.50
Minimum and maximum massage
pressure:
0.50 to 4.00 in 0.25 intervals