40
Interpreting test results
Welch Allyn MicroTymp3 portable tympanometric instrument
Influence of altitude
As the altitude above sea level increases, the admittance of a given volume of air also
increases. Therefore, equivalent ear canal volume (+200 Vea) overestimates actual ear
canal volume as noted in the table below. To estimate ear canal volume, subtract the
appropriate value in the table from the Vea Reading. Altitude can also affect MicroTest
Cavity results. See Functional Checks starting on
page 55
.
Influence of temperature
No corrections for temperature are required with this device.
Altit
u
de adj
u
stme
n
ts fo
r
Vea
r
eadi
n
gs
Altit
u
de
Adj
u
stme
n
t
fo
r
0.2 cc
Adj
u
stme
n
t
fo
r
0.5 cc
Adj
u
stme
n
t
fo
r
1.0 cc
Adj
u
stme
n
t
fo
r
1.5 cc
Adj
u
stme
n
t
fo
r
2.0 cc
Adj
u
stme
n
t
fo
r
2.4 cc
-1200 ft (-366 m)
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0 ft (0 m)
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.0 cc
1000 ft (305 m)
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.0 cc
2000 ft (610 m)
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.0 cc
0.1 cc
3000 ft (914 m)
0.0 cc
0.0 cc
0.1 cc
0.1 cc
0.1 cc
0.1 cc
4000 ft (1219 m)
0.0 cc
0.1 cc
0.1 cc
0.1 cc
0.1 cc
0.1 cc
5000 ft (1524 m)
0.0 cc
0.1 cc
0.1 cc
0.1 cc
0.1 cc
*
6000 ft (1829 m)
0.1 cc
0.1 cc
0.1 cc
0.2 cc
0.2 cc
*
7000 ft (2134 m)
0.1 cc
0.1 cc
0.1 cc
0.2 cc
0.2 cc
*
8000 ft (2438 m)
0.1 cc
0.1 cc
0.2 cc
0.2 cc
0.3 cc
*
9000 ft (2743 m)
0.1 cc
0.1 cc
0.2 cc
0.3 cc
0.3 cc
*
10000 ft (3048 m)
0.1 cc
0.2 cc
0.2 cc
0.3 cc
0.4 cc
*
* The MicroTymp3 will read “OPEN” when used on this ear canal volume at this altitude and will not run a test.
Summary of Contents for MicroTymp 3 Handle
Page 1: ...MicroTymp 3 portable tympanometric instrument Directions for use...
Page 6: ...vi Contents Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 16: ...16 Introduction Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 22: ...22 Completing a test Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 30: ...30 Understanding the LCD messages Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 48: ...48 Printing 226 Hz results Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 72: ...72 Specifications Welch Allyn MicroTymp3 portable tympanometric instrument...