Cleaning
Clean the handle
Clean the Handle by wiping it with a cloth that has been lightly dampened with 70%
Isopropyl alcohol. Make sure liquid does not seep into the instrument, especially in the
probe area.
Clean the printer/charger
Clean the Printer/Charger by wiping it with a cloth that has been lightly dampened with
70% Isopropyl alcohol. Make sure liquid does not seep into either the printer area or the
charging well.
Clean the probe tips
After each use, inspect small openings in the tip for debris, then remove tip from Handle.
If you observe no debris, wipe with a clean cloth and a 70% isopropyl alcohol solution.
Allow to air dry.
If you observe debris after use, rinse tip with warm tap water for one minute, then soak
the tip in 70% alcohol for 20 minutes. Rinse tip with tap water and air dry.