3
8
Interpreting test results
Welch Allyn MicroTymp3 portable tympanometric instrument
Normative values
The table below presents normative values for 1000-Hz tympanometry for NICU babies
and full-term 2- to 4-week-old babies. [Margolis, R.H., Bass-Ringdahl, S., Hanks, W.D.,
Holte, K., Zapala, D.A. Tympanometry in Newborn Infants - 1 KHz Norms. J. Amer. Acad.
Audiol., 14, 383-392, 2003.]
Interpreting 1000-Hz results
Pass
A normal tympanogram will yield a PASS result. The displayed numerical value represents
the compensated admittance.
Condition which causes a PASS message:
•
Peak Ya greater than or equal to 0.6 mmho
No
r
mative tympa
n
omet
r
ic val
u
es (1000 Hz)
Tympa
n
omet
r
ic Meas
ur
eme
n
t
NICU
90% Ra
n
ge
F
u
ll-te
r
m
90% Ra
n
ge
Peak Ya
(Negative Tail Compensation)
0.6 - 2.7 mmho
0.6 - 4.3 mmho
Tympanometric
Peak Pressure (TPP)
-93 - +53 daPa
-133 - +113 daPa
-400
-200
0
200
2
0.5
0
1.0
1
GR
A
GR
C
PRESSURE - daPa
Ya - mmho
+200 Vea - cc
Summary of Contents for MicroTymp 3 Handle
Page 1: ...MicroTymp 3 portable tympanometric instrument Directions for use...
Page 6: ...vi Contents Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 16: ...16 Introduction Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 22: ...22 Completing a test Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 30: ...30 Understanding the LCD messages Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 48: ...48 Printing 226 Hz results Welch Allyn MicroTymp3 portable tympanometric instrument...
Page 72: ...72 Specifications Welch Allyn MicroTymp3 portable tympanometric instrument...