Interpreting test results
Welch Allyn MicroTymp3 portable tympanometric instrument
Flat tympanograms
Flat tympanogram with normal ear canal volume
Condition which causes a flat tympanogram with normal ear canal volume:
•
Middle ear effusion
Flat tympanogram with increased ear canal volume
Conditions which cause flat graph and increased ear canal volume:
•
Patent tympanostomy tube
•
Perforated tympanic membrane
-400
-200
0
200
2
0.5
0
1.0
1
1.5
PRESSURE - daPa
Ya - mmho
Vea
+200
LEFT EAR
Note
This condition may produce a persistent OPEN message if measured Vea is
greater 2.5 cc. See
page 24
.
-400
-200
0
200
2
0.5
0
1.0
1
1.5
PRESSURE - daPa
Ya - mmho
Vea
+200
LEFT EAR