Catalog No. 29290
33
Rev 3 06/2020
4mA Adjustment:
•
The first menu outputs a constant 4mA while allowing for a small amount
of adjustment. This adjustment will allow the operator to make the
MTOL+ agree with a PLC or SCADA system. The adjustment limits are ±
200 counts or about ± 0.2 mA.
This setting will be slightly different on each instrument
as each MTOL+ will be factory set to 4.00mA. Press the
button when adjustments are complete to save this setting and
move on to the 20mA adjustment.
Setting the 4mA Levels
: Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press
the
button.
•
Press the
button to move to the next menu until
4MA
is displayed.
•
Adjust setting using the
and
buttons.
•
Press the
button when adjustments are complete to save this setting
and move on to the 20mA adjustment.
20mA Adjustment:
•
This menu operates similarly to the previous menu. This menu outputs a
constant 20 mA while allowing for a small amount of adjustment. The
adjustment limits are ± 1000 counts or about ± 1 mA.
•
This setting will be slightly different on each instrument as each MTOL+
will be factory set to 20.00mA.
Setting the 20mA Levels
: Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press
the
button.
•
Press the
button to move to the next menu until
20MA
is displayed.
•
Adjust setting using the
and
buttons.