Catalog No. 29290
38
Rev 3 06/2020
4.2.18 Update Software - Install Latest Software or Change Range
•
Enter the
CONFIG
mode by pressing the
and
buttons until the arrow
beside
CONFIG
is illuminated, then press the
button.
•
Press the
button to move to the next menu until
UPDT
is displayed,
then press the
button.
The software version can be determined by pressing and
holding down the
and
buttons for about 1 second.
The first number shown is the version and the second number
displayed is the model number. This can be performed anytime
during AUTO Mode.
•
Software updates can only be made if a USB flash drive is installed with a
newer version of software than the one currently installed in the MTOL+. If
either a flash drive is not installed or a newer version of the software is not
detected the following screen will appear.
•
Range changes can be made with either the current or later versions.
These other ranges were included on a flash drive included with the
instrument when new.
•
Once a flash drive with valid files are found, select the desired range to
update using the
and
buttons.
•
Press the
button to start the update.
•
While the system is updating, the word “busy” appears as it performs a
self-check of the software installation.