3.1 Site Selection
Choose a site that is accessible and well-lighted for use,
servicing, repair or replacement by authorized personnel.
•
MTOL+ should be located and mounted within 2-3 meters (6 to10 ft) of
the sampling point to ensure a quick response time.
•
Installation/mounting site must be isolated from vibration, and not
susceptible to freezing or temperature extremes.
•
The site must have a suitable vertical and flat surface that can support
the product’s weight and service operations 5kg (10lbs).
•
There should be 20 cm (8 in) free area above the instrument. This will
ensure enough room for calibration and cuvette maintenance.
•
The site must allow for mounting MTOL+ with the front display
positioned at eye level 1.4 to 2 meters (approximately 4.5 to 6.5 feet
from the floor).
•
MTOL+ is designed to have the Junction Box cradled under the upper
portion of the instrument – Select a site with enough space for proper
configuration when mounted.
3.2 Mounting
Described is for installation to a concrete surface. If needed
contact HF scientific for instruction to install on other
surfaces.
Tools required:
Appropriate screwdriver or driver gun.
Mounting Template – last page of this Manual.
Masking tape for attaching Mounting Template to the wall.
Any tools required by wall anchor manufacturer if used.
Hardware required:
4 Mounting screws for instrument enclosure - M6 (¼”). Either use concrete
screws or use a wall anchors & the supplied screws.
2 Mounting screws for terminal junction box - M4 (#8). Either use concrete
screws or use a wall anchors & the supplied screws.