Catalog No. 29290
24
Rev 3 06/2020
•
20MA
is displayed on the lower row of the LCD display. Select the
turbidity level to assign to the
20MA
using
the
and
buttons.
•
Once the desired level has been set, press the
button to accept the
setting and move to the next menu or the
button to exit and return to
AUTO
.
Error Level - Configuring the Error Level:
•
In case of an error in the MTOL+, the 4-20 mA reading can be used to
indicate a problem by sending the current to either
4.00
mA,
2.00
mA.
0.00
mA, or
OFF
.
•
The factory default setting is
OFF
and the 4-20mA is unaffected by any
error condition
.
•
To change the Error Level, enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press
the
button.
•
Press the
button to move to the next menu until
ERLV
is displayed.
•
Select the desired
ERLV
by using the
and
buttons then press the
button to accept the setting and move to the next menu or the
button to
exit and return to
AUTO
.
4.2.2 Baud Rate - Selecting the RS–485 Baud Rate
•
Select the desired baud rate (1200, 2400, 4800, 9600, or 19200) for the
operation of the I/O port.
To change the Baud Rate:
Enter the
CONFIG
mode by pressing the
and
buttons until the arrow beside
CONFIG
is illuminated, then press
the
button.
•
Press the
button to move to the next menu until
BAUD
is displayed.
•
Select the desired
BAUD
rate by pressing the
or
buttons to change
the displayed baud rate
.