Cleaning the source components
5-21
Cleaning the source components
Clean the sample cone and cone gas nozzle (see
) when these
conditions apply:
•
The sample cone and cone gas nozzle are visibly fouled.
•
LC and sample-related causes for decreased signal intensity have been
dismissed.
If cleaning the sample cone and cone gas nozzles fails to increase signal
sensitivity, also clean the extraction cone (see
).
If cleaning the extraction cone fails to increase signal sensitivity, clean the ion
block and isolation valve (see
If cleaning the ion block and isolation valve fails to increase signal sensitivity,
also clean the source T-Wave ion guide assembly (see
Cleaning the sampling cone assembly
You can remove the sampling cone assembly (comprising the sample cone,
O-ring, and cone gas nozzle) for cleaning without venting the instrument.
Removing the sampling cone assembly from the source
Required materials
Chemical-resistant, powder-free gloves
Summary of Contents for SYNAPT G2
Page 18: ...xviii Table of Contents...
Page 46: ...2 8 Starting Up and Shutting Down the Mass Spectrometer...
Page 66: ...3 20 Configuring the LockSpray Source...
Page 228: ...B 24 External Connections 7 Click Next Result The software installs 8 Click Finish...
Page 232: ...C 4 Materials of construction and compliant solvents...