B-8
External Connections
To connect the nitrogen gas supply
1. Connect one free end of the 6-mm PTFE tubing to the API gas in port on
the rear of the instrument.
Mass spectrometer gas and vacuum connectors
2. Attach a nitrogen regulator to the nitrogen supply.
Requirement:
The nitrogen must be dry and oil-free, with a purity of at
least 95%.
3. Connect the free end of the 6-mm PTFE tubing to the nitrogen regulator.
Warning:
The 6-mm (¼-in) PTFE tubing must not be cut to size; use
the complete 5-m (16-ft) length as supplied.
EXHAUST
PILOT VALVE
IMS GAS IN
1 bar
API GAS IN
7 bar
COLLISION
GAS IN
0.5 bar
HELIUM CELL
GAS IN
0.5 bar
PUMP RELAY
Roughing pump vacuum
connections
Nitrogen exhaust
Nitrogen inlet
Collision cell gas inlet
Pilot valve port
IMS gas inlet
Helium cell gas inlet