Replacing the LockSpray probe capillary
5-75
To remove the existing capillary
1. Prepare the instrument for working on the source (see
2. If an ESI probe is fitted to the source, remove it from the source (see
).
3. Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
4. Wait 10 minutes for the source ion block to cool.
5. Disconnect the "finger tight" nut on the inside of the source enclosure.
Warning:
The probe and source can be hot. To avoid burn injuries,
take great care while working with these components.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds, if
an ESI probe is fitted to the source, remove the probe before
continuing with this procedure.
"Finger
tight" nut
LockSpray sprayer
(behind baffle)