Removal and refitting of the source enclosure
5-9
4. Disconnect the source enclosure cables from the instrument’s
connectors.
5. Pull the source enclosure release (located at the bottom, right-hand side)
outwards, and swing open the enclosure.
6. Using two hands, grasp the source enclosure, and lift it vertically off the
two supporting studs on the source adaptor housing.
Warning:
The corona pin tip is sharp. To avoid puncture wounds,
take great care while working with the source enclosure open if a
corona pin is fitted.
Caution:
To avoid damaging the sample inlet when removing a
NanoLockSpray source enclosure, move the sprayer platform out
of the source enclosure before opening it (see
).
TP03164
Supporting stud
Source enclosure