Deploying the sprayer platform adjuster assembly
4-5
Deploying the sprayer platform adjuster assembly
To move the sprayer platform out of the source
1. Confirm that the sprayer’s safety cover is installed (see the figure on
).
2. Unscrew the thumbscrew on the front of the sprayer platform.
3. Pull out the side thumbscrew and withdraw the sprayer platform from
the source.
4. Release the side thumbscrew, locking the platform in the withdrawn
position.
NanoLockSpray source configuration
Sprayer type
Used for
Universal NanoFlow nebulizer
sprayer
Flow injection or for coupling to
nanoACQUITY UPLC with regulated
flow rates down to 100 nL/min.
Borosilicate glass capillary
NanoFlow (see
).
Extremely low flow rates (less than
100 nL/min).
NanoFlow capillary electrophoresis
(CE) sprayer. For details, see the
Capillary Electrophoresis/Capillary
Electrochromatography Sprayer
User's Guide
(part number 6666522).
Stable electrospray by means of a
make-up liquid at the CE capillary
tip. The make-up flow rate is less
than 1 µL/min.
Warning:
To avoid electrical shock, ensure the safety cover is in place
over the sprayer.
Summary of Contents for SYNAPT G2
Page 18: ...xviii Table of Contents...
Page 46: ...2 8 Starting Up and Shutting Down the Mass Spectrometer...
Page 66: ...3 20 Configuring the LockSpray Source...
Page 228: ...B 24 External Connections 7 Click Next Result The software installs 8 Click Finish...
Page 232: ...C 4 Materials of construction and compliant solvents...