Cleaning the source T-Wave ion guide assembly
5-51
Disassembling the T-Wave ion guide assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Jeweler’s screwdriver
To disassemble the T-Wave ion guide assembly
1. Using the jeweler’s screwdriver, remove the 3 screws securing the
locating ring to the T-Wave ion guide assembly.
2. Remove the metallized O-ring from the assembly.
Requirement:
If the metallized O-ring shows signs of deterioration or
damage, dispose of it in accordance with local environmental
regulations.
3. Remove the locating ring from the assembly.
Warning:
The T-Wave ion guide components can be
contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing
this procedure.
Warning:
The metallized O-ring can be contaminated with
biohazardous and/or toxic materials. Dispose of it according
to local environmental regulations.
TO
P
TO
P
Locating ring securing screw
Locating ring
Metallized O-ring
Aperture plate
Aperture plate
securing screw