![Waters ACQUITY System Manual Download Page 34](http://html2.mh-extra.com/html/waters/acquity/acquity_system-manual_3669408034.webp)
Long flats
Compression screw
2-piece stainless-steel ferrule
Tighten the fitting finger-tight plus an additional 3/4-turn using a 1/4-inch open-end wrench. For
detailed instructions about assembling new fittings, see
Tip:
To prevent band spreading, ensure that the tubing is fully bottomed in the fitting before
tightening the compression screw.
First use tightening
Re-installed
Long flats
Compression screw
2-piece stainless-steel ferrule
Tighten the fitting finger-tight plus as much as an additional 1/6-turn using a 1/4-inch open-end
wrench.
August 8, 2016, 715005049 Rev. C