Item
Specification
Leak control
Compartment drip tray with leak sensor installed under
compartment. Single exit drain manages leaks to waste.
Door open to heater cutoff
delay
1-minute maximum
Column tracking
eCord technology column information management tracks and
archives column usage history.
a. Operating with active pre-heating via the APH assembly
C.2.5 30-cm CHC performance specifications
Item
Specification
Settable temperature
range
4.0 to 65.0 °C, in 0.1 °C increments
Controllable temperature
range
(Ambient – 15 °C) to 65.0 °C
Temperature accuracy
Tested to ±0.8 °C between 20 and 50 °C (±1.0 °C for remaining
range)
Test qualification: The temperature measured by a NIST-traceable
probe located next to the measurement sensor must fall within this
specification.
Leak control
Compartment drip tray with additional condensate drain. Single exit
drain manages leaks to waste.
Door open to heater cutoff
delay
1-minute maximum
C.2.6 CM-A and CM-Aux performance specifications
Item
Specification
Settable temperature
range
4.0 to 90.0 °C, in 0.1 °C increments
Controllable
temperature range
(Ambient -25 °C) to 90.0 °C