C.2.3 CH-A performance specifications
Item
Specification
Settable temperature
range
20.0 to 90.0 °C, in 0.1 °C increments
Controllable temperature
range
(A5 °C) to 90.0 °C
Tested to ±0.5 °C between 20 and 50 °C (±1.0 °C for remaining
range)
Test qualification: The temperature measured by a NIST-traceable
probe located next to the measurement sensor must fall within this
specification.
Temperature stability
a
±0.3 °C at the sensor
Solvent conditioning
Active pre-heating as standard
Leak control
Compartment drip tray with leak sensor installed in sample manager
upper drip tray or under compartment (for extended configuration
only). Single exit drain manages leaks to waste.
Door open to heater
cutoff delay
1-minute maximum
Column tracking
eCord technology column information management tracks and
archives column usage history
a. Operating with active pre-heating via the APH assembly
C.2.4 CH-30A performance specifications
Item
Specification
Settable temperature
range
20.0 to 90.0 °C, in 0.1 °C increments
Controllable temperature
range
(A 5 °C) to 90.0 °C
Tested to ±0.5 °C between 20 and 50 °C (±1.0 °C for remaining
range)
Test qualification: The temperature measured by a NIST-traceable
probe located next to the measurement sensor must fall within this
specification.
Temperature stability
a
±0.3 °C at the sensor
Solvent conditioning
Active pre-heating
August 8, 2016, 715005049 Rev. C
Page 112