background image

SERVICE PARTS

 

All kits include applicable hardware and step-by-step instructions. 

KIT NAME

 

ORDER NO

 

DESCRIPTION 

 

Rubber Feet

 

2102273

 

Set of (4) replacement feet for legs.

 

Casters

 

2102275

 

Set of (4) replacement casters .

 

Short Caster Actuator

 

2102277

 

One short caster actuator for right rear or left front casters.

 

Long Caster Actuator

 

2102279

 

One long caster actuator for left rear or right front casters.

 

Lift Actuator

 

2102281

 

Set of (2) lift actuators.

 

Power Cord

 

2102283

 

Replacement power cord that plugs into wall receptacle.

 

Power Cord Hooks

 

2102285

 

Set of (2) Hooks on right side of LiftMate that secure the power cord.

 

E-Stop Interface 
Button

 

2102287

 

Red Button that attaches to the outside shroud. The emergency stop switch is inside the 
unit and is connected to the other side of this button. Kit includes (1) button.

 

E-Stop Switch

 

2102289

 

The switch that is mated to the E-Stop button. The kit includes (1) switch. The switch 
disconnects power to the lift actuators only.

 

Hand Pendant

 

2102291

 

Replacement hand pendant with strain relief and button graphics label attached.

 

Hand Pendant Hook

 

2102293

 

Hook for hanging hand pendant when not in use.

 

Back Cover

 

2102295

 

Sliding two piece cover that spans the gap between the platform and base.

 

Platform Mat

 

2102297

 

Slip resistant mat that covers the top of the platform.

 

Platform Bumper

 

2102299

 

Yellow rubber strip covering the front edge of the platform.

 

Right Outer Shroud

 

2102301

 

Right outside cover with power cord hooks and E-Stop button attached.

 

Left Outer Shroud

 

2102303

 

Left outside cover with E-Stop button attached.

 

Right Inner Shroud

 

2102305

 

Right inside cover.

 

Left Inner Shroud

 

2102307

 

Left inside cover.

 

Battery

 

2102309

 

Replaceable storage battery specifically designed for this unit.

 

Control Box

 

2102311

 

Control unit through which the power cord is connected and the battery is charged. The 
hand pendant input and the lift and caster actuator outputs are connected as well. Lift 
actuator synchronization and battery management is monitored through the control box.

 

E-Stop Wiring Harness

 

2102313

 

The wiring harness that connects to the E-Stop switches and allows power to operate the 
lift actuators when switches are in their normal state.

 

L Caster Actuator 
Harness

 

2102315

 

The wiring harness that brings power to the left caster actuators.

 

R Caster Actuator 
Harness

 

2102317

 

The wiring harness that brings power to the right caster actuators.

 

L Lift Actuator 
Harness

 

2102319

 

The wiring harness that brings power to the left lift actuator.

 

R Lift Actuator 
Harness

 

2102321

 

The wiring harness that brings power to the right lift actuator.

 

Battery Wiring 
Harness

 

2102323

 

The wiring harness that goes from the control box to a receptacle that the 
battery plugs into. Charge current to the battery and power from the battery 
travel through this harness.

 

Summary of Contents for Brewer Patient LiftMate

Page 1: ...ON MANUAL Component Identification 2 Important Information 3 Specifications 4 Safety Information 4 Electrical Requirements 5 Installation 6 Operation 8 Maintenance 11 Troubleshooting 11 Service Parts 14 Limited Warranty 15 ...

Page 2: ...COMPONENT IDENTIFICATION ...

Page 3: ...n either extend the casters to roll the LiftMate away from the table or LiftMate may be left in place depending on the type of exam and physician s preference When the examination is complete the LiftMate can be rolled back into position Once the casters have retracted and the LiftMate is resting on all four feet the platform can be raised to a height that is comfortable for the patient to step on...

Page 4: ... 18 minutes off motor run time Operating Conditions Temperature Range 65 F to 85 F Relative Humidity 10 to 90 Transportation and Storage Temperature Range 20 F to 150 F Humidity 10 to 90 SAFETY INFORMATION Safety Information The primary concern of The Brewer Company is that the equipment is operated and maintained with the safety of the patient and healthcare staff in mind To ensure safe and relia...

Page 5: ...ow more than one person on the LiftMate at a time Ensure that the back of the LiftMate is pushed as close to the exam table as possible Make sure that all exam table drawers stirrups attachments or other protrusions are pushed in to prevent damage Children Caregivers and Untrained Personnel Anyone not trained in the safe use of the LiftMate must not be allowed to operate it Ensure that children an...

Page 6: ...t any damage to the carrier immediately CAUTION To avoid damaging the LiftMate s plastic or painted surfaces DO NOT use a knife or other sharp object to open the packaging Lift only at points indicated in Figure 3 Do not lift at other points Do not plug LiftMate into receptacle prior to removal of all packaging materials WARNING The LiftMate weighs approximately 105 lbs Two or more people should a...

Page 7: ...ife to shorten if use continues all functions will stop The unit will beep and a steady indicator light on the hand pendant will appear if any hand pendant button is pushed Plug LiftMate into wall receptacle to enable functions and recharge battery 3 Always use a replacement battery from Brewer Never use any other battery which could cause damage to the unit or improper operation 4 When not using ...

Page 8: ...een the LiftMate and any other object Failure to do so could result in unwanted motion and possible injury to patient or operator or damage to LiftMate or exam table NOTE For optimum LiftMate performance allow the unit to reach room temperature before operating NOTE The LiftMate is not designed for continuous operation The control system includes a thermal overload safety switch to protect from ov...

Page 9: ...e of obstructions that may cause unit to tip Patient Handling with the LiftMate Positioning the LiftMate and raising the patient 1 Extend casters Figure 5 2 Position the back of LiftMate against foot end or side of exam table Figure 6 3 When in position allow casters to retract by releasing the button 4 Assist patient by having them walk forward onto LiftMate platform 5 Once patient is on LiftMate...

Page 10: ...a standing position on platform and grasp handrail 6 Lower platform all the way down Figure 9 and assist patient off of LiftMate Emergency Stops Emergency Stops are located on the outer right hand and outer left hand Shrouds Figure 10 In the unlikely event of unintended movement of the Platform depressing the Emergency Stop Figure 11 will cause all Platform movement to cease Have a qualified techn...

Page 11: ... water Avoid the use of harsh chemicals as they can permanently damage the finish To prevent mat from peeling do not allow it to remain wet TROUBLESHOOTING Casters All casters will not extend or retract This may indicate a low battery condition Pressing any function on the Hand Pendant will cause Low Battery light on Hand Pendant to illuminate and beeping sound to occur Charge battery Note pluggin...

Page 12: ...on the hand pendant simultaneously and hold for a 10 beep count emanating from the LiftMate When the10 beeps are complete a second cycle of beeps with a longer pause between them will emanate from the LiftMate It is during this second cycle of beeps that the LiftMate platform can be moved Press the Down button on the hand pendant to ensure that it is lowering correctly If it is lowering correctly ...

Page 13: ... protective surface that will not damage the finish Replace feet onto ends of legs by pressing all the way on and hold A slight bump will protrude from the centre of the foot Pushing this bump in will cause the conical shaped catch on the inside of the foot to hook into the receiving hole in the leg If there is difficulty getting the catch to hook a rounded device of sufficient diameter 3 16 or la...

Page 14: ... platform Platform Bumper 2102299 Yellow rubber strip covering the front edge of the platform Right Outer Shroud 2102301 Right outside cover with power cord hooks and E Stop button attached Left Outer Shroud 2102303 Left outside cover with E Stop button attached Right Inner Shroud 2102305 Right inside cover Left Inner Shroud 2102307 Left inside cover Battery 2102309 Replaceable storage battery spe...

Page 15: ...1800 337 711 03 9314 6300 Sales Enquiries sales warequip com au Contact Information Service and Spare Parts service warequip com au Customer Service customerservice warequip com au www warequip com au ...

Reviews: