LiftMate Operates Only When Plugged In
A. If the battery’s ability to hold a charge has decreased over time, the battery requires replacement.
B. If battery has suddenly failed without warning the battery may have become disconnected. Remove the left
outer cover to check battery connection at the plug in jack. If the connection is tight, and the battery is still
inoperable, replace the battery.
LiftMate Fails to Operate Even When Plugged In
Sudden failure of all functions is indicative of no power at the wall receptacle or failure of the control box. Plug a
lamp or other device known to operate into the receptacle to ensure power is available. If no power is available
plug LiftMate into wall receptacle proven to have power. Check to see if LiftMate functions resume operation. If
problem persists, contact Warehousing Equipment Customer Service
at 1800 337 711
for a replacement control
box.
Rubber Feet
The rubber feet are easily replaced. Lay LiftMate onto its back on a carpeted or protective surface that will not
damage the finish. Replace feet onto ends of legs by pressing all the way on and hold. A slight bump will
protrude from the centre of the foot. Pushing this bump in will cause the conical shaped catch on the inside of the
foot to hook into the receiving hole in the leg. If there is difficulty getting the catch to hook, a rounded device of
sufficient diameter (3/16”) or larger such as a ball end Allen wrench can be inserted into the hole and pushed in.
DO NOT USE A POINTED OR SHARP OBJECT SUCH AS A PHILLIPS OR FLAT BLADED
SCREWDRIVER.
See the Service Parts list for replacement kit number.
Platform Mat
If the platform mat has become excessively soiled, worn or is peeling, replace with the Platform Mat
Replacement Kit. See the Service Parts list for replacement kit number.
Platform Bumper
If the platform bumper has detached it can be pushed back onto the edge of the platform. If the bumper will not
stay attached replace with the Platform Bumper Kit. See the Service Parts list for replacement kit number.
Other Troubleshooting
For all other troubleshooting please contact Warehousing Equipment Customer Service
at 1800 337 711.