Weight of LiftMate............................................... 105 lbs.
Load Rating (maximum):
- Platform.............................................................. 500 lbs.
- Platform Range......................................... 1.0 to 10.0 in.
LiftMate Dimensions:
Platform Width (Inside)..................................... 21 3/8 in.
Platform Depth (Inside)..................................... 18 1/4 in.
Distance Between Inner Shrouds....................... 24 1/4 in.
Distance Between Platform Handrails (Inner Surfaces).28
in.
Width Between Outer Shrouds.......................... 34 3/4 in.
Base Width (Outer)............................................ 31 3/4 in.
Base Depth (Outer)............................................ 21 1/2 in.
Mat Dimensions
Width.................................................................. 19 1/2 in.
Width........................................................................ 17 in.
Caster Dimensions
Diameter..................................................................... 2 in.
Width.......................................................................... 2 in.
Electrical Specifications
115 VAC nominal, 60 HZ, 4 amps max.
SPECIFICATIONS
Power Cord..............................Extends 10 ft. from unit.
18 AWG / 3 conductor, SJT grey jacketed with hospital
plug
Type of protection against electrical shock
Class............................................. Class II Double
Insulated
Type of protection against electrical shock………. B
Applied Parts
Certifications
*
UL60601-1; IEC60601-1; CAN-CSA C22.2 No. 601.1
Battery Specifications
Type.............................................Sealed Lead Acid
Storage
Charging..........Battery is charged through Control Box
only
Temperature (Ambient Operating).......... 41°F to 104°F
Output...................................................... 24 V DC, 10A
Duty Cycle
..... 2 minutes on/18 minutes off (motor run
time)
Operating Conditions
Temperature Range..................................65°F to 85°F
Relative Humidity....................................10% to 90%
Transportation and Storage
Temperature Range.................................-20°F to 150°F
Humidity...................................................10% to 90%
SAFETY INFORMATION
Safety Information
The primary concern of The Brewer Company
is that the equipment is operated and maintained
with the safety of the patient and healthcare
staff in mind. To ensure safe and reliable
operation:
● Read and understand all instructions in this
manual before attempting to install or operate
the unit.
● Ensure that appropriate personnel are
informed on the manual contents. This is the
responsibility of the purchaser.
● Ensure that this manual is located near the
LiftMate, or if possible permanently affixed to
the unit.
Failure to follow instructions with this symbol
could result in serious personal injury and/or damage
to the unit.
DANGER indicates an imminently hazardous
situation which, if not avoided, will result in death or
serious injury.
WARNING indicates a potentially hazardous
situation which, if not avoided, could result in death
or serious injury.
CAUTION indicates an imminently or potentially
hazardous situation which, if not avoided, may
result in serious injury and/or equipment damage
Summary of Contents for Brewer Patient LiftMate