MAINTENANCE
Preventative Maintenance
Cleaning
CAUTION
Failure to perform periodic inspections of the
LiftMate could result in personal injury or
equipment damage.
CAUTION
The shroud material is resistant to most medicinal
type stains, but may be damaged by solvents and
dyes. Remove any spilled fluids from the shrouds,
painted surfaces or mat immediately.
● Periodically inspect the electrical cord to ensure it
is free of cuts or damage.
● Inspect the mechanical functions to ensure
satisfactory operation.
● Ensure that cord hooks are not loose. Remove
outer right shroud and tighten if necessary.
● Check shroud screws to make sure they are present
and tightened securely.
● Check platform bumper for damage, and to ensure
that it is secured to the edge of the platform. Reattach
or replace as necessary.
● Make sure the covers located at the back of the
platform when raised, are engaged with the guides,
and move freely when the platform is raised up and
down.
● Clean the LiftMate weekly, wiping the painted metal
and plastic surfaces with a clean, soft cloth. If desired,
mild cleaners may be used.
● Clean the mat immediately after contact with any
fluids or gels. Use a mild soap and rinse with clear
water. Avoid the use of harsh chemicals as they can
permanently damage the finish. To prevent mat from
peeling do not allow it to remain wet.
TROUBLESHOOTING
Casters
All casters will not extend or retract.
This may indicate a low battery condition. Pressing any function on the
Hand Pendant will cause “Low Battery” light on Hand Pendant to illuminate, and beeping sound to occur. Charge
battery. Note: plugging LiftMate into receptacle will resume all functions in a low battery condition. If this action
does not resume functionality contact Warehousing Equipment Customer Service at
1800 337 711.
One caster will not extend or retract.
Contact Warehousing Equipment Customer Service
at 1800 337 711.
Casters are difficult to roll.
Lay LiftMate onto its back on a carpeted or protective surface that will not damage
the finish. Check casters for wear, debris or damage. Clean or replace as needed. See the Service Parts list for
replacement kit number.
Casters causing marks on floor
. Over time casters can get embedded with dirt redepositing it onto flooring. Lay
LiftMate onto its back on a carpeted or protective surface that will not damage the finish. Clean casters with soap
and water. Use a fine stiff wire brush with just enough force to remove dirt (use only if necessary). Use this
opportunity to check casters for wear or damage. Replace as needed. See parts list for replacement kit number.
Summary of Contents for Brewer Patient LiftMate