background image

ELECTRICAL REQUIREMENTS

 

 Do not perform any procedures while 

patient is on the LiftMate. 

 Use 115 VAC, 60 Hz as noted on the rating 

label. Failure to do so may result in electrical 
shock to personnel and will result in damage 
to the LiftMate. 
● Do not use the LiftMate in an explosive or 
oxygen enriched atmosphere. Failure to do so 
may result in serious personal injury or death. 
● Charging the battery with any device other 
than through the Control Box on the unit may 
result in serious personal injury, damage to 
other LiftMate components, and/or premature 
failure of the battery.

 

WARNING

 

CAUTION

 

Do not use any power supply other than that listed on the 
rating label. Failure to do so may result in serious injury 
and/or equipment damage. 

●This product has been evaluated with respect to 

electrical shock, fire, and mechanical hazards only in 
accordance with UL60601-1; IEC60601-1; CAN/CSA 
C22.2 No. 601.1

 

Electromagnetic Interference 

This product is designed and built to minimize electromagnetic 
interference with other devices; however, if interference is 
noticed between another device and this product, remove the 
interfering device from the room or plug this product into an 
isolated circuit. 

WARNING

 

Never place any body parts 

between the Handrail Crossbrace and Shrouds, 
or under the 
front edge of the Platform. 

Both views show platform raised

 

WARNING

 Please observe the following 

warnings (continued). 

Do not allow more than one person on the LiftMate at a 
time. 
●Ensure that the back of the LiftMate is pushed as close to 
the exam table as possible. 
●●Make sure that all exam table drawers, stirrups, 
attachments or other protrusions are pushed in to prevent 
damage. 

Children, Caregivers and Untrained Personnel

●Anyone not trained in the safe use of the LiftMate must not 
be allowed to operate it. 
●Ensure that children and others are clear of the LiftMate at 
all times whether it is being operated or not. 
●Never allow the patient to operate the LiftMate at any time. 

Special Patient Situations

●The LiftMate must not be used by patients to pull 
themselves up out of a wheelchair or walker. 
●Do not use the LiftMate for patients that are unable to 
support their own weight. 
●Never exceed LiftMate’s maximum capacity of 500 lbs. 

General

: Please observe the following consideration. 

●LiftMate must not be used to transport anything including 
patients. 
● If battery fails while casters are up or down, plug into 
receptacle to recharge. 

Patient and Operator: 

Ensure that loose 

clothing, buttons, jewelry or other such items that 
can snag onto the LiftMate are kept clear of catch 
points 

Patient

●Never put a patient on the LiftMate while 
casters are extended. 
●Patient should only get on or off of the LiftMate 
when all four base feet are firmly on the ground. 
●Do not allow patients to pull themselves up by 
using any part of the LiftMate. 

●Patient must stand in the centre of the platform during the 
lift procedure. Never lean forward, backward or sideways 
beyond the area indicated by the dashed line. 

WARNING

 Please observe the 

following warnings. 

Summary of Contents for Brewer Patient LiftMate

Page 1: ...ON MANUAL Component Identification 2 Important Information 3 Specifications 4 Safety Information 4 Electrical Requirements 5 Installation 6 Operation 8 Maintenance 11 Troubleshooting 11 Service Parts 14 Limited Warranty 15 ...

Page 2: ...COMPONENT IDENTIFICATION ...

Page 3: ...n either extend the casters to roll the LiftMate away from the table or LiftMate may be left in place depending on the type of exam and physician s preference When the examination is complete the LiftMate can be rolled back into position Once the casters have retracted and the LiftMate is resting on all four feet the platform can be raised to a height that is comfortable for the patient to step on...

Page 4: ... 18 minutes off motor run time Operating Conditions Temperature Range 65 F to 85 F Relative Humidity 10 to 90 Transportation and Storage Temperature Range 20 F to 150 F Humidity 10 to 90 SAFETY INFORMATION Safety Information The primary concern of The Brewer Company is that the equipment is operated and maintained with the safety of the patient and healthcare staff in mind To ensure safe and relia...

Page 5: ...ow more than one person on the LiftMate at a time Ensure that the back of the LiftMate is pushed as close to the exam table as possible Make sure that all exam table drawers stirrups attachments or other protrusions are pushed in to prevent damage Children Caregivers and Untrained Personnel Anyone not trained in the safe use of the LiftMate must not be allowed to operate it Ensure that children an...

Page 6: ...t any damage to the carrier immediately CAUTION To avoid damaging the LiftMate s plastic or painted surfaces DO NOT use a knife or other sharp object to open the packaging Lift only at points indicated in Figure 3 Do not lift at other points Do not plug LiftMate into receptacle prior to removal of all packaging materials WARNING The LiftMate weighs approximately 105 lbs Two or more people should a...

Page 7: ...ife to shorten if use continues all functions will stop The unit will beep and a steady indicator light on the hand pendant will appear if any hand pendant button is pushed Plug LiftMate into wall receptacle to enable functions and recharge battery 3 Always use a replacement battery from Brewer Never use any other battery which could cause damage to the unit or improper operation 4 When not using ...

Page 8: ...een the LiftMate and any other object Failure to do so could result in unwanted motion and possible injury to patient or operator or damage to LiftMate or exam table NOTE For optimum LiftMate performance allow the unit to reach room temperature before operating NOTE The LiftMate is not designed for continuous operation The control system includes a thermal overload safety switch to protect from ov...

Page 9: ...e of obstructions that may cause unit to tip Patient Handling with the LiftMate Positioning the LiftMate and raising the patient 1 Extend casters Figure 5 2 Position the back of LiftMate against foot end or side of exam table Figure 6 3 When in position allow casters to retract by releasing the button 4 Assist patient by having them walk forward onto LiftMate platform 5 Once patient is on LiftMate...

Page 10: ...a standing position on platform and grasp handrail 6 Lower platform all the way down Figure 9 and assist patient off of LiftMate Emergency Stops Emergency Stops are located on the outer right hand and outer left hand Shrouds Figure 10 In the unlikely event of unintended movement of the Platform depressing the Emergency Stop Figure 11 will cause all Platform movement to cease Have a qualified techn...

Page 11: ... water Avoid the use of harsh chemicals as they can permanently damage the finish To prevent mat from peeling do not allow it to remain wet TROUBLESHOOTING Casters All casters will not extend or retract This may indicate a low battery condition Pressing any function on the Hand Pendant will cause Low Battery light on Hand Pendant to illuminate and beeping sound to occur Charge battery Note pluggin...

Page 12: ...on the hand pendant simultaneously and hold for a 10 beep count emanating from the LiftMate When the10 beeps are complete a second cycle of beeps with a longer pause between them will emanate from the LiftMate It is during this second cycle of beeps that the LiftMate platform can be moved Press the Down button on the hand pendant to ensure that it is lowering correctly If it is lowering correctly ...

Page 13: ... protective surface that will not damage the finish Replace feet onto ends of legs by pressing all the way on and hold A slight bump will protrude from the centre of the foot Pushing this bump in will cause the conical shaped catch on the inside of the foot to hook into the receiving hole in the leg If there is difficulty getting the catch to hook a rounded device of sufficient diameter 3 16 or la...

Page 14: ... platform Platform Bumper 2102299 Yellow rubber strip covering the front edge of the platform Right Outer Shroud 2102301 Right outside cover with power cord hooks and E Stop button attached Left Outer Shroud 2102303 Left outside cover with E Stop button attached Right Inner Shroud 2102305 Right inside cover Left Inner Shroud 2102307 Left inside cover Battery 2102309 Replaceable storage battery spe...

Page 15: ...1800 337 711 03 9314 6300 Sales Enquiries sales warequip com au Contact Information Service and Spare Parts service warequip com au Customer Service customerservice warequip com au www warequip com au ...

Reviews: