background image

Step

Description

Image

1

Slowly opens the water network tap and insert the
Osmo mains cable into the power socket.

Note:

make sure there is no leakage and there are no

traces of water on the bottom of the equipment.

Do not operate the equipment without having
opened the water supply first.
The pump can suffer irreparable damage if run dry.

2

Let the water come out of the “DRAIN” tube until it is
clean and free from black bits and other residuals.

3

Open the storage tank tap.

6. Priming

20

5 ÷ 15 min

Summary of Contents for Osmo

Page 1: ...Osmo ENG Rev 00 Instructions for use...

Page 2: ...Contents of the packaging 9 4 Conditions of use and period of use 11 5 Installation 12 6 Priming 20 7 Use and operation 21 8 Maintenance 22 9 Water Block safety valve 27 10 Spare parts 29 11 Troubles...

Page 3: ...ent of risk of damage to property W H symbols Consult the instructions for use High voltage Do not open or tamper with the equipment Live electric parts inside the unit Do not dispose of with normal w...

Page 4: ...ranty covers replacement or repair of the components recognized as being unsuitable due to manufacturing defects inclusive of the necessary labor The above is not applicable to faults caused by incorr...

Page 5: ...al absence of chemical products The Osmo can eliminate from water all impurities and harmful pollutants Treated water is the ideal solution for professional use and delivery since the exclusive filtra...

Page 6: ...General information and Safety advice 6 This equipment requires earthing BEFORE STARTING UP THE EQUIPMENT Check that the equipment has not been damaged Do not use products which have been damaged dur...

Page 7: ...ing to the instructions in this manual or is not properly maintained the manufacturer cannot be held responsible for any fault damage or malfunctioning of the equipment If the tap water shows high qua...

Page 8: ...ther The permitted temperature is 5 C 35 C Even if carefully packaged and protected the system must be considered and handled as fragile material On receipt check the external conditions of the packag...

Page 9: ...Open the package and check that the contents correspond to the delivery note and to the list in the next page there is no evidence of any damage In the event of damage or missing items inform the car...

Page 10: ...nal connecting tubes support bracket kit optional 2 7 litre storage tank with a tap 3 Water Block valve with filter and setting key 4 Opening closing tool for osmotic vessel cap 5 Dowels and screws fo...

Page 11: ...lacement and flush filter cartridge for 5 minutes See label on the equipment and instructions for use Resin filter cartridge life Max 6 months or exhaustion Replacement See instructions for use Osmoti...

Page 12: ...m WB Valve The safety valve Water Block prevents a water flow greater than the set value which is the flow requested by the appliance connected This prevents big leaks in case a vessel or a tube gets...

Page 13: ...unted installation Install the osmotic group using the 4 supplied dowels and screws OR Installation with brackets optional Install the brackets on the osmotic group using the 4 screws and nuts Install...

Page 14: ...8 mm to the Water Block valve outlet Attention seal the thread with PTFE tape Do not use glue nor paste as a sealant 4 Remove the red cap from the IN fitting of the Osmo and tighten the quick connecto...

Page 15: ...e the tube out of the fitting press the fitting ring with your fingers and pull out the tube Note To allow the complete installation of the Osmo the white tube needs to be cut into several pieces 6 Co...

Page 16: ...om the fitting TANK of the Osmo in the storage tank tap and tighten it with the ring nut 9 Connect the white tube in the fitting OUT of the Osmo This is a quick connector fitting insert the tube until...

Page 17: ...he water dispenser in the other free fitting of the T connection Note tighten the tubes connected to the T connection with the ring nuts 12 Connect the white tube in the fitting DRAIN of the Osmo This...

Page 18: ...Installation 18 Example of installation Connection to Water dispenser Tap optional...

Page 19: ...Installation 19 EXAMPLES OF USAGE Description Image Manual use with water dispenser for manual filling of the sterilizer tank and washing of instruments Manual usage with tap optional...

Page 20: ...ge and there are no traces of water on the bottom of the equipment Do not operate the equipment without having opened the water supply first The pump can suffer irreparable damage if run dry 2 Let the...

Page 21: ...O AVOID WATER CONTAMINATION 7 Use and operation 21 In accordance with subsection 4 5 of EN1717 in order to avoid water quality degradation and bacterial growth the entire system must be drained before...

Page 22: ...year in case a sterilizer is attached to the system this could be done during the scheduled maintenance visits for the sterilizer It is further recommended to check all related devices fitted to the s...

Page 23: ...over 15 S after the filter cartridge maintenance kit has been replaced also replace the osmotic membrane All the filter cartridges can vary slightly which in combination with unfavourable operational...

Page 24: ...hausted filter cartridge counter clockwise and remove it Note during this operation water will inevitably pour out of the vessel 5 Insert the new filter cartridge and turn it clockwise to the end unti...

Page 25: ...e electrical power supply 2 Close the water inlet tap and the storage tank tap 3 If present release the water pressure with the use of the water dispenser or tap if connected optional 4 Using the appr...

Page 26: ...he new membrane completely making sure that the two o rings are positioned correctly Tighten the osmotic vessel cap by turning it clockwise with the appropriate tool Note check carefully that there ar...

Page 27: ...0 N cm2 Max controllable volume about 50 lt Min operating capacity 2 lt min Max operating capacity 30 lt min Max water temperature 70 C USER INSTRUCTIONS Water Block should be installed between the ta...

Page 28: ...f and carry out the necessary repair work Remove the supply hose C from the Water Block and press the coloured plunger H which is located inside the lower housing of the device itself fig 1 It may be...

Page 29: ...lter cartridges A813030X Complete maintenance kit resin carbon filter cartridges osmotic membrane A813053X No 2 quick connector fittings tube 8 mm A813031X Pump 230 V A813032X Pump 115 V A813033X Sole...

Page 30: ...Resin filter cartridge head A813043X Carbon filter cartridge head A813042X Osmotic vessel A813046X L shaped quick fitting 8 mm for black tube A813047X L shaped quick fitting 6 mm for white tube A8130...

Page 31: ...ption Code Drawing Image Pressure switch A813052X Water dispenser kit A812018X Water tap kit optional A813044X Support bracket kit optional A813045X Pressure reducer x 3 bar optional A813025X Spare pa...

Page 32: ...ower supply water inlet presence Water inlet problem Water Block operated Set the Water Block properly The motor pump starts by itself at regular intervals Leak on the supply circuit downstream of the...

Page 33: ...ed power 230 V 138 W Max absorbed power 115 V 138 W Class of insulation III Min max supply pressure 0 5 5 bar Usage environment Indoor Weight with filter cartridges empty storage tank 13 kg Weight of...

Page 34: ...ized and managed by the supplier The user shall contact the supplier and follow the method that has been implemented for disposing of the equipment at the end of its life The correct disposal techniqu...

Page 35: ...Notes...

Page 36: ...Osmo ENG Instructions for Use Rev 00 Subject to alterations 08 01 2021 W H Sterilization S r l Italy I 24060 Brusaporto BG via Bolgara 2 t 39 035 66 63 000 f 39 035 50 96 988 wh com...

Reviews: