5. Installation
12
PRELIMINARY INSTALLATION INSTRUCTIONS
Wall-mounted installation
The water demineralizer should be fitted in a vertical position either beneath a sink or in a cabinet.
For mounting, use the supplied dowels and screws.
Install on the wall leaving a space of at least 10 cm from the floor or the bottom surface so that the filter cartridge can be dismantled.
Overall dimensions
Power supply
The equipment is supplied with 230 V ac or 115 V ac and the maximum power absorbed is 138 W.
Hydraulic installation
Drain
Install a drain nearby to dispose the waste water discarded by the osmotic membrane (max 1.5 m).
WB Valve
The safety valve (Water Block
®
) prevents a water flow greater than the set value, which is the flow requested by the appliance connected.
This prevents big leaks in case a vessel or a tube gets broken. The Water Block is also fitted with a special check, preventing the return of
water into the hydraulic system and possible consequent contamination.
For this purpose, the WB valve must be installed in a vertical position (refer to section 9).
Charge
The installation of the product with inlet water pressure above 5 bars, can drastically reduce the lifespan of the equipment.
In case the inlet water pressure is more than 5 bars, install a suitable pressure reducer.
230 mm
3
5
0
m
m
520 mm
3
7
0
m
m
When the storage tank is full of
water, it weights approximately
10 kg.
Check that the work surface
can withstand this weight.