55
vwr.com I Instruction manual VWR® G Water System
AC-Pack Not Detected
·
Reinstall the AC Pack.
·
Check the RFID chip.
·
Contact a VWR’s professional engineer.
P-Pack Not Detected
·
Reinstall the P Pack.
·
Check the RFID chip.
·
Contact a VWR’s professional engineer.
U-Pack Not Detected
·
Reinstall the U Pack.
·
Check the RFID chip.
·
Contact a VWR’s professional engineer.
Feed Temperature>45
o
C
·
Contact a VWR’s professional engineer.
Feed Temperature<5
o
C
·
Contact a VWR’s professional engineer.
RO Product Temperature>45
o
C
·
Contact a VWR’s professional engineer.
RO Product Temperature<5
o
C
·
Contact a VWR’s professional engineer.
EDI Product Temperature>45
o
C
·
Contact a VWR’s professional engineer.
EDI Product Temperature<5
o
C
·
Contact a VWR’s professional engineer.
UP Product Temperature>45
o
C
·
Contact a VWR’s professional engineer.
UP Product Temperature<5
o
C
·
Contact a VWR’s professional engineer.
TOC Sensor Temperature>45
o
C
·
Contact a VWR’s professional engineer.
TOC Sensor Temperature<5
o
C
·
Contact a VWR’s professional engineer.
TOC Feed Resistivity<15MΩ.cm
·
Contact a VWR’s professional engineer.
Cannot dispense water from dispenser. It
shows “In Use”
·
Another dispenser connected to the same
systems is being used. System allows one
dispenser working at a time. Wait till the
other dispenser stopped dispensing.
System Leakage or Tank Overflow
·
The system detects a leakage.
·
Shut off power and water supply.
·
Remove system side panels, turn on water
supply and check leaking points.
·
Reconnect or replace leaking parts.