vwr.com I Instruction manual VWR® G Water System
a) Power off the VWR main system.
b) Screw off the nut on the back of the controller. Remove the controller and carefully remove the
used lamp from the chamber. Unplug the UV lamp base.
c) Wear gloves included in the new UV lamp package. Avoid direct skin contact with the quartz
glass of the UV lamp. Install the new UV lamp to the lamp base, and then gently insert the
new UV lamp completely into the chamber.
d) Screw the sealing cap on the tank cap, and then screw the nut to fix the controller on the tank
cap.
e) Turn on the power and operate the system. Scan the RFID tag for the UV sanitization module
on the RFID reader position ( ) to register it.
Nut
Sealing Cap
Tank Cap
Controller
Support