61
vwr.com I Instruction manual VWR® G Water System
a)
Turn off the power.
b)
Remove the left side panel to find the UV lamps.
c)
Loosen the UV lamp chamber screws. As shown above, the blue circles indicate the UP water
UV lamp chamber screws, and the yellow circles indicate the EDI water UV lamp chamber
screws.
d)
Tilt the upper end of the UV lamp chamber out.
e)
Wear gloves included in the new UV lamp package. Avoid direct skin contact with the quartz
glass of the UV lamp. Remove the UV lamp cover to expose the UV lamp. Unplug the UV
lamp from its power cord, and then carefully remove the old UV lamp from the chamber.
UP UV lamp
EDI UV lamp