65
vwr.com I Instruction manual VWR® G Water System
c) Using a Phillips screwdriver, remove the 5 screws in the red circle on the white back panel.
d)
Find the pipe filters shown as the red circle in the left side picture below, remove the blue
buckles shown as the yellow circle in the right side picture below.
e) Remove the pipe filters and install a new one.
Note: Please confirm each filter needs to be placed in the right direction.
5.4.8 Cleaning the Inlet filter
a) Switch the system to Standby mode, Tap the dispense button
on the dispenser to
dispense water. Gas inside the system will be released out of the system. Close the Inlet Valve.
b) Remove the system right side panel.
c) Turn the lever on the cartridge adapter to right to the unlock position, pull the cartridge
downwards to disengage the cartridge from its adapter. Remove all the cartridges.
d) Using a Phillips screwdriver, remove the 5 screws on the white back panel.
e)
Find the inlet filters shown as the red circle in the left side picture below, remove the blue