62
vwr.com I Instruction manual VWR® G Water System
f)
Carefully insert the new UV lamp into its chamber. When about 2/3 of the lamp is inserted,
hold the UV lamp and connect it to the ballast cable connector (4-pin connector) as shown in
the picture, and then gently insert the new UV lamp completely into the chamber.
g)
Cap the UV chamber with the black mask.
h)
Install the UV lamp chamber back to the original position.
i)
Turn on the system power. Scan the RFID chip of the UV lamp on the system. Tap
Install
on
the pop-up dialog
to reset the corresponding UV lamps.
5.4.3 Replacing the Tank UV Lamp
Warning!
Ultraviolet (UV) radiation is harmful to the eyes and skin. Do not observe the lamp
directly when it is illuminated. This system is equipped with a lamp cover to prevent
UV light leakage. This cover must be on ALL TIMEs when a UV lamp is installed.
Caution!
Keep the UV lamp straight in and out of the stainless steel chamber during its
installation to avoid any action that could cause the lamp to break.