Step 5 Connect the signal cables to ASM port and OVER FLOW port on the system
respectively.
2.4.6 Installing the Water Tubing
2.4.6.1 Connecting the Feed and Drain Tubing
d) Remove the stopper on the tap water inlet of the system. Take out the 8 mm PE tubing from
the accessories bag and cut three pieces of the tubing into proper lengths for feed and drains.
e) Insert one end of the first 8 mm PE tubing into the water inlet of the system (Mark as
Tap
Water In
). Connect the other end to the tap water outlet. Gently pull the tubing outward to
check if the connection is secure.
f)
Insert one end of the second 8 mm PE tubing into the
RO Drain
of the system. Put the other
end to the drain.
g) Insert one end of the third 8 mm PE tubing into the
EDI Drain
of the system. Put the other end
to the drain.
HP Dispenser
UP Dispenser
Tap Water
EDI Drain
RO
Drain