Chapter 6 Troubleshooting and Repair
M-000-00030
6-7
Analog Output Calibration
To check the 4
–
20 mA circuit, connect a DVM in series with the output loop. Select
zero or full scale (from the second level of the hidden Diagnostics menu) and then
actuate the
ENTER
key. This action will cause the meter to output its 4 mA or 20 mA
condition. If the DVM indicates a current greater than ± 0.006 mA from 4 or 20 mA,
adjust the setting up or down until the output is calibrated.
Note: these settings are not for adjusting the output zero and span to match a flow
range, that function is located in the Output Menu.
Summary of Contents for SonoPro S34 Series
Page 34: ...Series S34 Instruction Manual Chapter 3 Operation 3 5 M 000 00030 Output Menu...
Page 69: ...Series S34 Instruction Manual Appendix A Specifications A 2 M 000 00030...
Page 73: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 2 M 000 00030 Pipe Chart...
Page 74: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 3 M 000 00030...
Page 75: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 4 M 000 00030...
Page 76: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 5 M 000 00030...
Page 77: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 6 M 000 00030...
Page 78: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 7 M 000 00030...
Page 79: ...Series S34 Instruction Manual Appendix C Sound Speed and Pipe Data C 8 M 000 00030...