![viZaar VUCAM AM+ Operating Instructions Manual Download Page 37](http://html.mh-extra.com/html/vizaar/vucam-am/vucam-am_operating-instructions-manual_1041497037.webp)
check before each operation that the protection ring
is screwed onto the probe. Proceed very carefully
when mounting and screwing. never use force. If in
doubt, ask our service department for advice.
to attach the protection ring to the probe (double
thread), follow the steps below:
1. Mount the protection ring onto the first thread of
the head of the probe.
2. Slightly press the protection ring to the second
thread. Mount the protection ring onto the second
thread. to remove the protection ring, please pro-
ceed in reverse order.
3 sET-UP AnD CoMMissioning
never use the probe without the pro-
tection ring. otherwise, the probe may
be damaged.
The tip adapters may only be changed
by trained or experienced personnel.
The thread on both the probe and the
adapters is very fine due to their small
dimensions. To avoid damage, pay at-
tention to small impurities in the
thread!
To FiT (CloCkwisE)
To REMovE (AnTiCloCkwisE)
To FiT (CloCkwisE)
To REMovE (AnTiCloCkwisE)
1
2
3
37
* only with VucaM
®
Xo/XF-Models
3.6
ConnECTion oF
PRoTECTion Ring*