44
12.2
Output to Peripherals
Our standard peripheral units can be connected to the balance.*1
These peripheral units include: CSP-160, CSP-240
12.2.1
Connector pin numbers and functions
Pin No.
Signal
Name
Input/Output
Function & Remarks
1
EXT.TARE
Input
Tare setting from an
external device *2
2
-
-
-
3
-
-
-
4
TXD
Output
Transmitting data
5
GND
-
Signal ground
6
-
-
-
7
-
-
-
8
-
-
-
*1 To connect a peripheral unit to the balance, use a connection cable supplied with the unit.
*2 You can set a tare range or adjust the zero point from an external device by connecting a
contact or a transistor switch between pin 1 and pin 5. In this case, allow at least 400 ms
for connection (ON) time (Maximum voltage when the balance is turned OFF: 15 V, sink
current when it is turned ON: 20 mA).
Caution:
Be sure to disconnect the AC adapter from the power outlet before plugging or unplugging the
connector.
DIN8P connector for output to
peripheral devices
1
2
3
4
5
6
7
8