34
The Span test will be performed to find an error of measure without making Span adjustment.
Please use it to investigate the error of measure.
9.3
Span Test with Built-in Weight
* Only supported by the HJR-KCE series
1 Set Function 1.
Set Function 1 with nothing put on the pan.
Select span test with built-in weight in Function 1
[
].
2 Start a span test.
Press the Cal key.
The built-in weight is enabled, and automatically starts a span
test.
The display indicates [
], [
], [
],
[
], and an error in sequence. When you press a key,
the balance goes back into measurement mode.
* An error displayed after [
] is calculated as shown below:
Error = True value – Current weight
A positive value displayed as an error means that the balance indicates a weight lighter than the
actual weight.
9.4
Span Test with External Weight
1 Set Function 1.
Set Function 1 with nothing put on the pan.
Select span test with external weight in Function 1
[
].
2 Start a span test.
Press the Cal key.
The display first indicates [
], and then [
],
and the balance starts testing the zero point. (If the display
indicates [
], press the Function key).
When the display indicates [
], put a weight on the
pan. (If the display indicates [
], press the Function
key).
The balance starts a weighing capacity point test.
The display first indicates [
] and then an error.
When you press a key, the balance goes back into
measurement mode.
CAL