55
13 Use Printers
13.1
Setting up the Printer
(1) Use CSP-160 or CSP-240 with the balance.
(2) Set proper print functions (print control) with the balance referring to the instruction manual for
your printer.
The factory default of our printer is manual printing (printer control).
(3) Make the baud rate and other settings compatible between the balance and the printer.
13.2
Saving CAL and span test results
1 Set Function 1.
Set the ISO/GLP/GMP compatible item to
[
] and set the GLP compatible item
for measurement data to [
].
2 Performs span adjustment/span test.
During output, it indicates [
] and it looks
like the balance is frozen. However, this is a normal
operation.
Wait until the output operation is
completed.
13.3
Output of Measurement Data in Compliance with ISO/GLP/GMP
1 Set Function 1.
Set the ISO/GLP/GMP compatible item to
[
] and set the GLP compatible item
for measurement data to [
].
2 Header output
Press and hold the Print key for a while.
The display indicates [
] and a header is
output.
3 Measurement data output
Measurement data can be output at any time during
measurements.
4 Footer output
Press and hold the Print key for a while after the
measurement is completed.
The display indicates [
] and a footer is
output.
Caution
(1) Upon printing calibration/span test results or ISO/GLP/GMP compatible outputs, date and time
data is also printed.
Check the date and time set in the balance before printing.
(Refer to Section 10: Date and
Time Setup in page 40).
(2) Do not press any keys on the printer during printing.
* Refer to Section 14 "Output in Compliance with ISO/GLP/GMP" described from page 56 for
sample prints.
長押し
Continuous press
長押し
Continuous press