FVAm™ Series Benchtop MPO Autofocus Microscope User Guide
22132675 Rev. 101, Standard, 2-20
Getting started
Setting up Pass/Fail tests
Setting up Pass/Fail tests using FiberChekPRO
Step 1 Connect the FVAm microscope to FiberChekPRO.
Step 2 Click the
Inspection
tab to access options for the FVAm microscope.
Step 3 Select options from the following drop-down lists (see
“Setting up Pass/Fail tests” on page 2-19
):
▪ OPTICAL SETTING
▪ CONNECTOR
▪ PROFILE
Note:
Only options available on the microscope are present in the drop-down lists.
— End —
Setting up Pass/Fail tests using the touchscreen
The
Connector Setup
option on the
Settings
menu functions like an application Wizard, stepping you through the
process of selecting applicable predefined options for the type of fiber connector or transceiver module with MT
ferrule to be tested.
•
On the
Settings
menu, select
CONNECTOR SETUP
, and then select an option from the
SELECT PROFILE,
SELECT CONNECTOR
, and
SELECT TIP
sub-menus as they appear.
The following images show the sequence for selecting predefined options for testing MPO fiber connectors.
Note:
The
AUTO
option automatically determines the appropriate optical settings for the fiber connector being
tested.
— End —