FVAm™ Series Benchtop MPO Autofocus Microscope User Guide
22132675 Rev. 101, Standard, 2-19
Getting started
Setting up Pass/Fail tests
Setting up Pass/Fail tests
The FVAm microscope provides factory-set options you can use to set up Pass/Fail tests. Factory-set options cover
the following parameters:
•
Analysis profiles, which define Pass/Fail thresholds for fiber types
•
Connector definitions, which define fiber configuration
•
Optical settings, which define lighting and sizing calibration applied during inspection
To set up the FVAm microscope for performing Pass/Fail tests, you can use either FiberChekPRO or, if available,
the touchscreen and onboard controls to select the appropriate analysis profile for the fiber connectors or
transceiver modules to be tested, and then the appropriate connector definition.
Note:
FiberChekPRO provides tools that let you create custom options. For information, about creating
customized analysis profiles and optical settings, refer to the
FiberChekPRO Automated Fiber Inspection
and Analysis Software User Manual
. For information about creating customized connector definitions,
“Managing custom connector definitions” on page 3-9
.
Table 2-4: Factory-set options for Pass/Fail tests
Analysis Profile
Application
Connector
Definition
Optical Setting
Ribbon, SM APC
(IEC-61300-3-35-Ed. 2.0)
Ribbon cables with single-mode angle polished
connectors.
• MPO 12 x 1
• MPO 12 x 2
• MPO 16 x 1
• MPO 16 x 2
• MPO 8 x 1
• Ribbon Tips - 1
• Ribbon Tips - 2
• Ribbon Tips - 3
Ribbon, MM
(IEC-61300-3-35-Ed. 2.0)
Ribbon cables with single-mode (SM) ultra-physical
contact (UPC) connectors.