DGT / DGT60 / DGTQ / DGTP
20
WEi.PoS
REMOTE SCALE WEIGHT POSITION
In this step one sets the position of the first character of the weight value, in the string transmitted by the
remote scale, knowing that the first character on the left of the string has the 00 position.
A possible character sign is also part of the weight value.
For example, if the received string is
sppp CR + LF:
Received string
s
p
p
p
p
p
p
p
u
u
CR LF
Position of the
character
00
01
02
03
04
05
06
07
08
09
10
11
Therefore one should set the 00 value.
It is possible to set up to 2 characters (from 0 to 39).
(!) 00
WEi.LEn
LENGTH OF THE WEIGHT DATA STRING
In this step one enters the number of digits (from 1 to 39) which make up the weight value, including the
sign and the non significant digits (for example, if the transmitted string
spppppppuu + CR + LF
, in
which
s
is the weight sign,
ppppppp
is the weight value,
uu
is the unit of measure, one should set the
08 value).
Str.LEn
LENGTH OF THE COMPLETE STRING
In this step one enters the number of digits (from 0 to 39) which make up the entire string transmitted
from the serial line scale, less the terminator character (for example, if the transmitted string is
sppp CR + LF
, in which
s
is the weight sign,
ppppppp
is the weight value,
uu
is the unit of
measure, one should set the value 11).
(!) 01
dECi.
REMOTE SCALE DIVISIONS
Enter the number of scale’s decimals (from 0 to 5).
(!) 0
The two following parameters allow to generate the weight stability communicated by the remote scale
and allow to manage the stability LEDs on the indicator:
StAb
NUMBER OF READINGS PER STABILITY
Enter the number of consecutive readings which the indicator must consider to obtain the stability (2
characters, from 0 to 20).
(!) 03
StA.int
WEIGHT DIFFERENCE PER STABILITY
Enter the maximum value (2 characters, from 0 to 20) of the difference between the weights of the
consecutive readings, set in the previous step.
If the weight difference between the readings is equal or less than the set value, the weight is
considered to be stable (stability LED off), otherwise the weight is considered to be unstable (stability
LED on).
(!) 02
Pr1577
Reception of the “rEPE.6” string.
(§)
The parameter is displayed only if one has selected
the rEPE functioning mode in the
FModE
>>
FunCt
step.
(!) ondE
NOTES:
-
If the functioning mode selected in the
FModE
>>
FunCt
step is different than rEPE, by selecting the
REPE6 or REPE4 protocol the serial output is automatically set at 4800, N-8-1.
By selecting the Pr1577 protocol the serial output is automatically set at 4800, E-7-1.
In any case it’s possible to set it differently.
Summary of Contents for DGT
Page 65: ...DGT DGT60 DGTQ DGTP 65 11 CONNECTION SCHEMES 11 1 DGT4 DGT60 DGT4 AN MOTHERBOARD I O BOOT J 1...
Page 68: ...DGT DGT60 DGTQ DGTP 68 11 2 DGT4 PB MOTHERBOARD I O BOOT J 1...
Page 71: ...DGT DGT60 DGTQ DGTP 71 11 3 DGTQ DGTQ AN BOARDS...
Page 72: ...DGT DGT60 DGTQ DGTP 72 I O BOOT J1 J2 CAL...
Page 76: ...DGT DGT60 DGTQ DGTP 76 11 4 DGTQ PB BOARDS...
Page 77: ...DGT DGT60 DGTQ DGTP 77...
Page 81: ...DGT DGT60 DGTQ DGTP 81 11 5 DGTP DGTP AN DGTP PB MOTHERBOARD...