DGT / DGT60 / DGTQ / DGTP
24
o
AoMin
MINIMUM VALUE
Setting of the minimum value of the analogue output.
o
SiGn
SIGN (§)
PoSit
Functioning with positive weight.
nEGAt
Functioning with negative weight.
(§)
The parameter is displayed only if one has selected the operating mode on the Net Weight.
See the 7 "ANALOGUE OUTPUT (DGT4 AN and DGTQ AN)" section for configuring it.
•
dEFAu
INITIALIZATION OF THE INSTRUMENT
Through this step one can initialize the instrument with the subsequent activation of the default parameters. By
pressing PRINT, a confirmation message ("dFLt?) will appear: confirm again with PRINT or exit with any other
key.
NOTE: The initialization of the instrument causes a cancellation of the present calibration and the
activation of the default parameters. In any case if one exits the setup environment WITHOUT
CONFIRMING the modification made, all the parameters of the last saving made will remain (including
the calibration).
In case of approved instrument, the default DOES NOT HAVE EFFECT on the metrological parameters
(those marked with (*) ).
diAG DIAGNOSTICS MENU
It is a
submenu inside which it is possible to check the software components and the scale hardware.
PrG.Ver
CHECKING THE SOFTWARE VERSION
By pressing
the instrument shows the software version in the XX.YY.ZZ. format.
diV.int
CALIBRATION INTERNAL DIVISIONS(§)
By pressing
the instrument shows the calibration internal divisions.
(§)
The parameter is not displayed if one has selected the rEPE or MAStr functioning mode,
FmodE
>>
FunCt
parameter.
AdC.uV
MICRO VOLTS (§)
By pressing PRINT the instrument shows the micro volts relative to the weight on the scale.
With the ZERO and TARE keys it is possible to switch the visualisation of the microvolts for each configured
channel of the scale. In the "DEP.CH." mode it is also possible to view the sum of the microvolts of the configured
channels ("SuM" appears for a few instants).
NOTE:
The maximum voltage that the instrument accepts in input is 30 mV (30000
µ
V); the weighing system is
powered by the indicator at 5 Vdc.
A correct operation will have a value less than 30000 with a weight of full scale capacity on the weighing system.
(§)
The parameter is not displayed if one has selected the rEPE or MAStr functioning mode,
FmodE
>>
FunCt
parameter.
AdC.MVV
MILLIVOLTS (§)
By pressing
the instrument displays the millivolts relative to the weight on the scale.
Through the
ZERO
and
TARE
keys it’s possible to switch the display of the millivolts fot each configured channel
of the scale. With the
MODE
key one can execute the scale zero, while with
one cancels.
AdC.Pnt
CONVERTER POINTS
By pressing
the instrument shows the A/D converter points relative to the weight on the scale.
With the ZERO and TARE keys it is possible to switch the display of the A/D converter points for each scale’s
configured channgel. In the "DEP.CH." mode, it is also possible to view the sum of the microvolts of the
configured channels ("SuM" appears for a few instants).
(§)
The parameter is not displayed if one has selected the rEPE or MAStr functioning mode,
FmodE
>>
FunCt
parameter.
Summary of Contents for DGT
Page 65: ...DGT DGT60 DGTQ DGTP 65 11 CONNECTION SCHEMES 11 1 DGT4 DGT60 DGT4 AN MOTHERBOARD I O BOOT J 1...
Page 68: ...DGT DGT60 DGTQ DGTP 68 11 2 DGT4 PB MOTHERBOARD I O BOOT J 1...
Page 71: ...DGT DGT60 DGTQ DGTP 71 11 3 DGTQ DGTQ AN BOARDS...
Page 72: ...DGT DGT60 DGTQ DGTP 72 I O BOOT J1 J2 CAL...
Page 76: ...DGT DGT60 DGTQ DGTP 76 11 4 DGTQ PB BOARDS...
Page 77: ...DGT DGT60 DGTQ DGTP 77...
Page 81: ...DGT DGT60 DGTQ DGTP 81 11 5 DGTP DGTP AN DGTP PB MOTHERBOARD...