DGT / DGT60 / DGTQ / DGTP
38
Reading command of the extended weight with Average Piece Weight (APW)
[CC]REXTA<CR LF>
Instrument response in the "IND.CH." or in the "DEP.CH" mode: EXTENDED STRING (see paragraph 6.6.2).
Instrument response in the "TRANSM." mode: MULTISCALE STRING (see paragraph 6.6.3).
NOTES:
- If the instrument is in the "IND.CH." mode (scale with independent channels) or in the "DEP.CH" mode (scale with
dependent channels and digitally equalized) the weight value is read relative to the active channel; to read the
values of the other channels (if configured) one has to switch each time onto the desired channel (
see also CGCH
channel switching command
).
- If the instrument is in the "TRANSM." (scale with independent channels) mode it is possible to read simultaneously
the values for all the configured channels
Extended weight read command
[CC]REXD<CR LF>
Instrument response in the "IND.CH." or in the "DEP.CH" mode: EXTENDED STRING (see paragraph 6.6.2).
Instrument response in the "TRANSM." mode: MULTISCALE STRING (see paragraph 6.6.3).
Weight read command
[CC]READ<CR LF>
Instrument response: STANDARD STRING (see paragraph 6.6.1).
Weight reading command with sensitivity times 10
[CC]GR10<CR LF>
Instrument response: STANDARD STRING (see paragraph 6.6.1).
Reading command of micro volts relative to the weight
[CC]MVOL<CR LF>
Instrument response in "IND.CH." mode: STANDARD STRING (see paragraph 6.6.1).
Response of the instrument in "TRANSM" and "DEP.CH" mode: MULTISCALE STRING (See section 6.6.3).
NOTES:
- If the instrument is in the "IND.CH." mode (scale with independent channels) the voltage value is read relative to the
active channel; to read the values of the other channels (if configured) one has to switch each time onto the desired
channel (
see also CGCH channel switching command
).
- If the instrument is in the "TRANSM." (scale with independent channels) or in the "DEP.CH" mode (scale with
dependent channels and digitally equalized) mode it is possible to read simultaneously the values for all the
configured channels
Reading command of converter points relative to the weight
[CC]RAZF<CR LF>
Instrument response in "IND.CH." mode: STANDARD STRING (see paragraph 6.6.1).
Response of the instrument in "TRANSM" and "DEP.CH" mode: MULTISCALE STRING (See section 6.6.3).
NOTES:
- If the instrument is in the ("IND.CH.") mode (scale with independent channels) the voltage value is read relative to
the active channel; to read the values of the other channels (if configured) one has to switch each time onto the
desired channel (
see also CGCH channel switching command
).
- If the instrument is in the "TRANSM." (scale with independent channels) or in the "DEP.CH" mode (scale with
dependent channels and digitally equalized) mode it is possible to read simultaneously the values for all the
configured channels
Tare command
[CC]TARE<CR LF> or [CC]T<CR LF> (short command).
Instrument response: [CC]OK<CR LF> if the command has been RECEIVED; the instrument’s response does not
mean necessarily that the instrument executes the tare.
Summary of Contents for DGT
Page 65: ...DGT DGT60 DGTQ DGTP 65 11 CONNECTION SCHEMES 11 1 DGT4 DGT60 DGT4 AN MOTHERBOARD I O BOOT J 1...
Page 68: ...DGT DGT60 DGTQ DGTP 68 11 2 DGT4 PB MOTHERBOARD I O BOOT J 1...
Page 71: ...DGT DGT60 DGTQ DGTP 71 11 3 DGTQ DGTQ AN BOARDS...
Page 72: ...DGT DGT60 DGTQ DGTP 72 I O BOOT J1 J2 CAL...
Page 76: ...DGT DGT60 DGTQ DGTP 76 11 4 DGTQ PB BOARDS...
Page 77: ...DGT DGT60 DGTQ DGTP 77...
Page 81: ...DGT DGT60 DGTQ DGTP 81 11 5 DGTP DGTP AN DGTP PB MOTHERBOARD...