INSPECTION
In principle we recommend one inspection every twelve months, and a minimum of one before usage
is resumed. All of the following checks must be performed and documented by authorized persons:
•
Check the frame parts and the hinged tubes for plastic deformation, tears and impaired
functioning
•
Visually check for damage to the paintwork (danger of corrosion)
•
Check the operation of the wheels (free running, level rolling, axle play, tyres, profile, condition
of the rims, air pressure (in the case of pneumatic tyres), etc.)
•
Check the solidity and seating of all screws
•
Verify the amount of grease on the metal joints of movable parts
•
The condition and security of the guides and axles of the steering wheels (castor nuts)
•
Visually check all plastic parts for tears and brittle spots
•
Check the functioning of the armrests and leg supports (locking, load, deformation, wear and
tear caused by loads)
•
Check the operation of other detachable parts (e.g. anti-tipping device, personal safety belt,
headrest, fitting of back/seat, container for walking stick, etc.)
•
Verify completeness of the delivery condition
•
Is the instruction manual available?
The service must only be signed off in the service plan if a minimum of all the above-mentioned
aspects have been checked. If your specialist dealer is incapable of performing the required services,
please contact the manufacturer. We shall gladly refer you to authorized specialist dealers in your
area.
The manufacturer is not liable for damage caused by lacking or improper service.
DISINFECTION
Only a hygiene technician or someone appointed by him can disinfect your wheelchair and this should
be done every time before its use is resumed. All parts of the wheelchair can be treated with a
disinfectant cloth. In principle all surfaces of a system or a product must be disinfected before handing
it over to a new user, at least once a year (e.g. when it is inspected) or when infections are detected in
the user. Then the measures of the Federal Epidemic Law must be applied.
We recommend the following disinfectants for scrubbing (based on the list provided by the