Operations & Maintenance Manual: Maintenance & Safety
0900‑5005 REV‑00
PROCEDURE 2. RUN A SELF‑TEST
The BladderScan i10 self‑test function completes self‑diagnostic testing on the following:
• Internal memory (and external devices, if any)
• Hardware components, including the console, installed battery, probe, and printer, if attached
• BladderScan i10 software
1. On the Home screen, tap the
Settings
icon, and then tap
Self‑Test
. The Settings screen opens to the
Self‑Test tab, where the system completes the self‑test and displays the results.
2. Scroll through the on‑screen test results by vertically swiping with your finger on the touchscreen.
3. If the screen indicates any failed tests or abnormal results, contact Verathon Customer Care or your local
representative.
4. If you want to print the self‑test results, tap
.
5. When you are finished viewing the test results, tap
Back
. The console returns to the Home screen.