1. Go to
>Utilities >Settings >More >High Precision Clock >Atomic Clock.
2. Set
Discipline Source = Ext. 1PPS (RJ11)
.
3. Verify that
1PPS Signal Health = Valid.
4. Set
Discipline Profile = Custom.
5. Set
Time Constant (s) = 600.
6. Set
Discipline Threshold (ns) = 20.
7. Verify that
Disciplining Status = Acquiring.
8. Tap on the
Phase Graph
to monitor the relative phase error, until it becomes a horizontal line at about 0ns.
9. Close the Phase Graph and wait until the
Disciplining Status = Locked.
At this point the test set’s internal “Atomic 10MHz” frequency and “Atomic 1PPS” timing reference signals can be considered highly
accurate and stable.