test set will automatically freeze the image when the image comes into focus. Once the image is frozen, users can tap on the image
to save it.
Analysis ON/OFF:
Turns the optional auto-analysis feature ON and OFF. If turned ON, the test set automatically analyzes the
image in real time. If the markers get in the way of seeing the picture details, users can turn this feature OFF and perform manual
analysis.
Dots/Rectangles:
They are two ways of highlighting the different impurities found in the face of the connector. The dots tightly
contour each individual anomaly, capturing its area. In some cases users may want to see impurities or scratches that are otherwise
covered by the dots, so rectangles are used in this case to highlight the anomaly and still keep them visible. The selection between
dots or rectangles does not affect the area calculation of the Pass/Fail results; they are just the operator’s choice.
Save:
The Save button appears after freezing an image. It performs the same action as tapping on the screen to save the current
image. Use the >Utilities >Files function to View the report, export to PDF format and copy to USB.
MMF/SMF:
Selects the type of connector/fiber to be analyzed and loads the correct Pass/Fail mask. Please note that the label in the
button DOES NOT indicate the current mask type. It indicates that the mask would be changed to if activated.
Shake ON/OFF:
Turning the Shake ON may help in situations when the image is not very stable, such as inspecting a female
connector or bulkhead.
7.5.1.8 Captured Files Tab
This section of the user interface lists all the connectors’ end face images that have been captured with the fiber inspection
microscope. Users can add “Before” and “After” identifiers or reminders when comparing images before or after cleaning or
polishing.
TX300s_Platform_Manual_RevB00
Page 79 of 104
Summary of Contents for TX300s
Page 1: ......
Page 15: ...Go back to TOC 3 5 Home Screen Go back to TOC TX300s_Platform_Manual_RevB00 Page 15 of 104...
Page 29: ...Time Zone Setup Go back to TOC TX300s_Platform_Manual_RevB00 Page 29 of 104...
Page 75: ...ARP Wiz Result Go back to TOC TX300s_Platform_Manual_RevB00 Page 75 of 104...
Page 100: ...GNSS Save Results Go back to TOC TX300s_Platform_Manual_RevB00 Page 100 of 104...