22
03-914786-00:3
NEEDLE WASH
(YES/NO)
Enter 1 to select needle wash.
Positions 93-96 are now used as containers for washing solvent.
After every sample vial the needle and connecting tubing is
flushed with solvent from one of the “wash vials” (alternating
between 93 and 96).
Enter 0 if no needle wash is wanted (position 93-96 can now be
used for sample vials).
AUXILIARIES (YES/NO)
Enter 1 if time programmable auxiliary output(s) are to be used.
Parameters will appear in the RUN PARAMETER mode.
Enter 0 if no auxiliaries are to be used.
RUN Parameters
With the ProStar 400 in READY status, press PROG/END to
select RUN PARAMETER mode and to enable programming at
the same time.
The following parameters can be programmed (use STEP key to
go to next parameter):
FIRST VIAL
Enter position number of the first vial of the sample series in the
sample tray.
1-96 without needle wash
1-92 with needle wash
1-48 with LSV
LAST VIAL
Enter position number of the last vial of the sample series in the
tray. (See
FIRST VIAL
for valid numbers.)
INJ/VIAL
Up to 9 injections can be taken from the same vial. Enter number
0-9.
PRE FLUSH VOLUME
Parameter shows up for flush on a
volume
basis only; enter
volume (of sample) which is to be used to flush needle and
connecting tubing prior to filling the loop: 30-999 µL.
FLUSH TIME
Parameter shows for flush on a
time
basis only. Enter duration of
headspace pressure for loopfilling: 1-999 sec.
ANALYSIS TIME
Enter the time interval needed before the next injection takes
place: 0-99'59". Time countdown starts at moment of injection.
Summary of Contents for ProStar 400