26
03-914786-00:3
A “next injection” input signal will start the RUN.
A blinking
R
(first position of the lower display line) indicates that
the ProStar 400 is under remote control. All other relevant RUN
information will be displayed as usual.
When all programmed vials have been processed, the display
will show:
0 REMOTE CONTROL
F,0 TO EXIT
Next injection or next vial inputs will no longer have any effect.
Exit REMOTE CONTROL mode (F,0) to return to READY status.
PC Workstation Interface
Up to four ProStar 400 AutoSamplers can be simultaneously
controlled from the Varian Star Chromatography Workstation.
Control requires purchase of the Star LC Control Software Kit.
Contact your local Varian representative for more information or
to order communication kits.
Figure 7 shows the connection cabling and the kit part numbers
for the Star Workstation control of ProStar 400 AutoSamplers.
Complete installation instructions are included with these kits.