28
03-914786-00:3
MEMORY ERROR
ALL VALUES DEFAULT
Press any key; if the ProStar 400 goes to READY status, all
parameters are set to default values. Reprogram RUN and
TEMP.CONTROL parameters and check SYSTEM SETTINGS.
If message remains, call Varian.
TRAY ROTATION
FAILURE
Tray movement is obstructed; check/remove obstruction. Press
any key to continue; if message remains: call Varian.
NEEDLE MOVEMENT
FAILURE
Needle arm movement is obstructed; check/remove obstruction.
Press any key to continue; if message remains, call Varian.
DISPENSER FAILURE
Call Varian.
PROGRAM VIALS
SYSTEM SETTING STD/LSV was changed: FIRST VIAL and
LAST VIAL are reset to 0; reprogram (RUN PARAMETER
mode). Press any key to continue.
NEEDLE WASH was switched ON (SYSTEM SETTINGS) with
FIRST or LAST vial programmed on WASH positions. FIRST
VIAL and LAST VIAL are reset to 0; reprogram (RUN
PARAMETER mode). Press any key to continue.
ERROR.(.............)
An erratic value has been entered; limits for this parameter are
shown in parenthesis on display. Enter correct value.
After Use
If the ProStar 400 AutoSampler is not to be used for more than
two days, flush needle and valve with water and isopropanol
(can be placed as last vials of a sample series) and loosen pump
tubing cassette screw one turn. Never leave the AutoSampler
out of use with a buffer or salt solution in the tubing; crystals may
block the flow path and may damage the rotor seal of the
injection valve. Power can be switched OFF since all parameter
values are saved in memory (long-life non-chargeable battery
back-up). Only sample tray cooling is changed to OFF after
power UP.